Ultrasonic Thickness Gauges - DM5E
|
|
The DM5E series wall thickness gages enables you to accomplish a large number of demanding tasks, especially with remaining wall thickness measurements on components subjected to corrosion and erosion.
The DM5E Series wall thickess gages are smart, simple-to-use ultrasonic thickness gages that offer big ultrasonic performance in a small lightweight package. Compact size and ergonomic shape provide comfortable one-hand use.
|
Instrument:
|
|
| Operating Principle |
Ultrasonic pulse-echo method |
| Measurement Modes |
Depends on Probe and Material, 0.60 mm to 508 mm (0.025” to 20.00”) in IP to 1st BW Measurement Mode, 2.00 mm to 127.0 mm
(0.079” to 5.00”) in Dual-Multi Measurement Mode, the Coating Thickness Range Shall be 0.3 mm to 2.50 mm (0” to 0.098”). |
| Measuring Resolution |
0.01 mm Default - Selectable 0.01, 0.1 mm (0.001” Default - Selectable 0.001, 0.01”) |
| Material Velocity Range |
0.508 to 18.699 mm/msec (0.0200 to 0.7362”/msec) |
| Material Velocity Resolution |
1 m/s (0.0001”/msec) |
| Units |
Inch or Millimeter |
| Calibration |
One-Point Cal, On-Block and Off-Block, Two-Point Cal |
| Pulser |
Excitation Pulse: Spike Pulser
Voltage: 120 V into 50 ohm Load, Using 20 MHz Oscilloscope |
| Receiver |
Bandwidth 500KHz to 12 MHz @ -3 dB
Gain Automatic Gain Control |
| Display Type |
High-Resolution Graphical LCD, 64 x 128 Pixels, 53.0 mm x 27.0 mm with Backlight and Adjustable Contrast |
| Update Rate |
4 Hz or 8 Hz, User Selectable, 24 Hz Scan Mode Capture Rate |
| Thickness Value Display |
NORMAL View Mode 5 Digit, 10.6 mm (0.4”) High
B-SCAN View Mode 5 Digit, 2.55 mm (0.1”) High |
| Display of Last Reading |
Solid Filled or Hollow Digits Indicate Coupled or Uncoupled Condition |
| Setups |
9 Standard Setups for Probes |
| Alarm Settings |
Minimum and Maximum Alarms, Range of 0.25 mm to 508 mm, 0. (0.010” to 20.00”)
Reading Alternates Between Solid and Hollow When Alarms Are Enabled and Violated |
| Power Requirements |
2 “AA” Size Batteries |
Battery Life/Operating
Time |
Approximately 60 Hours |
| Instrument Shut-0ff |
Selectable ALWAYS ON or AUTO OFF After 5, 10, 15, 30 Minutes of Inactivity |
| Language |
Selectable English, German, French, Spanish, Italian, Russian, Japanese and Chinese |
| I/O Connectors |
Transducer Dual Lemo 00 (coax)
Mini-USB Mini-USB to PC |
| Temperature |
Operating -10°C to +50°C (+10°F to +120°F)
Storage -20°C to +60°C (-10°F to +140°F) |
| Weight |
223 g (0.597 lb) Including Batteries |
| Size |
138 mm x 32 mm x 75 mm |
| Shock |
IEC 68-2-27 Ea, as per Mil Std 810C Method 516.2 Procedure I with a 15g 11ms Impulse Half Sinusoidal Wave Applied 6 Times per Axis |
| Sealing |
IEC529 / IP54, Dust Proof/Dripping Water Proof as per IEC 529 Specifications for IP54 Classification |
Data Record Option Features:
|
|
| Capacity |
50,000 Readings |
| File Structures |
Grid File |
| Number of Rows |
1 to 50,000 |
| Number of Columns |
1 to 223 |
| File Naming |
Up to 24 Character Alphanumeric Name |
| Optional Software |
UltraMATE and UltraMATE Lite |
Probe/Transducers:
|
|
|